Comparison of the Accuracy of Clinical Measurements and Cone-Beam Computed Tomography for Assessing Peri-Implant Buccal Bone Thickness Around Titanium and Zirconia Implants

Autores
Bono, Alejandra; Grosso Rossi, Sacha; Ibañez, Juan Carlos
Año de publicación
2026
Idioma
inglés
Tipo de recurso
artículo
Estado
versión publicada
Descripción
Fil: Bono, Alejandra, Liliana. Universidad Nacional de Córdoba. Facultad de Odontología. Departamento de patología oral; Argentina.
Fil: Ibañez, Juan Carlos. Universidad Católica de Córdoba. Facultad de Ciencias de la Salud. Especialización en Implantología Oral; Argentina.
Fil: Grosso Rossi, Sacha. Universidad Nacional de Córdoba. Facultad de Odontología. Cátedra Cirugía II; Argentina.
Dental implants have been an effective solution for restoring oral function and aesthetics in patients with tooth loss. Currently, technological advances have introduced new materials, such as zirconia, which have sparked significant interest in the scientific community. This research aims to compare zirconia and titanium implants and highlight the importance of cone beam computed tomography (CBCT) in the diagnosis of bone complications such as buccal bone dehiscence or fenestrations. Dental implants are a growing part of the prosthetic field. Zirconia has gained popularity as an implant material due to its color and advantageous biological potential. Although it demonstrates soft tissue adhesion and osseointegration potential comparable to commonly used titanium, buccal bone thickness is typically assessed during implant installation or second-stage surgery in submerged implant protocols, but is rarely evaluated thereafter, as conventional oral radiographic diagnosis only assesses interdental marginal bone levels. This factor, however, can have important implications for the long-term stability of peri-implant hard and soft tissues. Computed tomography (CBCT) has significantly expanded the use of CT imaging in implant dentistry due to its potentially reduced radiation dose and cost. However, CBCT images of peri-implant bone can be affected by artifacts that can generate incorrect bone gray values and, consequently, erroneous reconstructions. This study aims to evaluate the accuracy of measuring peri-implant buccal bone using cone-beam computed tomography devices on zirconium implants compared to titanium implants. Objective: To evaluate the accuracy of cone-beam computed tomography on zirconium and titanium implants for measuring buccal peri-implant bone. Implants were placed at different distances from the cortical bone in one prepared porcine rib. The rib will be scanned in a reproducible position using a CT scanner. An examiner will take four measurements on the eight implants: cortical bone thickness from the outer surface of the implant threads at 0, 3, 6, and 9 mm from the top of the implant.
info:eu-repo/semantics/publishedVersion
Fil: Bono, Alejandra, Liliana. Universidad Nacional de Córdoba. Facultad de Odontología. Departamento de patología oral; Argentina.
Fil: Ibañez, Juan Carlos. Universidad Católica de Córdoba. Facultad de Ciencias de la Salud. Especialización en Implantología Oral; Argentina.
Fil: Grosso Rossi, Sacha. Universidad Nacional de Córdoba. Facultad de Odontología. Cátedra Cirugía II; Argentina.
Materia
Dental Implants
Zirconia
Cbct
Vestibular Table
Nivel de accesibilidad
acceso abierto
Condiciones de uso
Repositorio
Repositorio Digital Universitario (UNC)
Institución
Universidad Nacional de Córdoba
OAI Identificador
oai:rdu.unc.edu.ar:11086/562556

id RDUUNC_352b8755b62990e62b6a07db821d4a57
oai_identifier_str oai:rdu.unc.edu.ar:11086/562556
network_acronym_str RDUUNC
repository_id_str 2572
network_name_str Repositorio Digital Universitario (UNC)
spelling Comparison of the Accuracy of Clinical Measurements and Cone-Beam Computed Tomography for Assessing Peri-Implant Buccal Bone Thickness Around Titanium and Zirconia ImplantsBono, AlejandraGrosso Rossi, SachaIbañez, Juan CarlosDental ImplantsZirconiaCbctVestibular TableFil: Bono, Alejandra, Liliana. Universidad Nacional de Córdoba. Facultad de Odontología. Departamento de patología oral; Argentina.Fil: Ibañez, Juan Carlos. Universidad Católica de Córdoba. Facultad de Ciencias de la Salud. Especialización en Implantología Oral; Argentina.Fil: Grosso Rossi, Sacha. Universidad Nacional de Córdoba. Facultad de Odontología. Cátedra Cirugía II; Argentina.Dental implants have been an effective solution for restoring oral function and aesthetics in patients with tooth loss. Currently, technological advances have introduced new materials, such as zirconia, which have sparked significant interest in the scientific community. This research aims to compare zirconia and titanium implants and highlight the importance of cone beam computed tomography (CBCT) in the diagnosis of bone complications such as buccal bone dehiscence or fenestrations. Dental implants are a growing part of the prosthetic field. Zirconia has gained popularity as an implant material due to its color and advantageous biological potential. Although it demonstrates soft tissue adhesion and osseointegration potential comparable to commonly used titanium, buccal bone thickness is typically assessed during implant installation or second-stage surgery in submerged implant protocols, but is rarely evaluated thereafter, as conventional oral radiographic diagnosis only assesses interdental marginal bone levels. This factor, however, can have important implications for the long-term stability of peri-implant hard and soft tissues. Computed tomography (CBCT) has significantly expanded the use of CT imaging in implant dentistry due to its potentially reduced radiation dose and cost. However, CBCT images of peri-implant bone can be affected by artifacts that can generate incorrect bone gray values and, consequently, erroneous reconstructions. This study aims to evaluate the accuracy of measuring peri-implant buccal bone using cone-beam computed tomography devices on zirconium implants compared to titanium implants. Objective: To evaluate the accuracy of cone-beam computed tomography on zirconium and titanium implants for measuring buccal peri-implant bone. Implants were placed at different distances from the cortical bone in one prepared porcine rib. The rib will be scanned in a reproducible position using a CT scanner. An examiner will take four measurements on the eight implants: cortical bone thickness from the outer surface of the implant threads at 0, 3, 6, and 9 mm from the top of the implant.info:eu-repo/semantics/publishedVersionFil: Bono, Alejandra, Liliana. Universidad Nacional de Córdoba. Facultad de Odontología. Departamento de patología oral; Argentina.Fil: Ibañez, Juan Carlos. Universidad Católica de Córdoba. Facultad de Ciencias de la Salud. Especialización en Implantología Oral; Argentina.Fil: Grosso Rossi, Sacha. Universidad Nacional de Córdoba. Facultad de Odontología. Cátedra Cirugía II; Argentina.https://orcid.org/0000-0003-3684-1947https://orcid.org/0009-0004-5292-1714https://orcid.org/0000-0001-6857-48182026-08-01info:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articlehttp://purl.org/coar/resource_type/c_6501info:ar-repo/semantics/articuloapplication/pdfGrosso Rossi S., Ibanez J.C. & Bono A. (2026). Comparison of the Accuracy of Clinical Measurements and Cone-Beam Computed Tomography for Assessing Peri-Implant Buccal Bone Thickness Around Titanium and Zirconia Implants. J Oral Dent Health Res, 8(3): 211. https://doi.org/10.36266/JODHR/2112694-6211http://hdl.handle.net/11086/56255610.36266/JODHR/211enginfo:eu-repo/semantics/openAccessreponame:Repositorio Digital Universitario (UNC)instname:Universidad Nacional de Córdobainstacron:UNC2026-10-01T10:23:28Zoai:rdu.unc.edu.ar:11086/562556Institucionalhttps://rdu.unc.edu.ar/Universidad públicaNo correspondehttp://rdu.unc.edu.ar/oai/snrdoca.unc@gmail.comArgentinaNo correspondeNo correspondeNo correspondeopendoar:25722026-10-01 10:23:30.263Repositorio Digital Universitario (UNC) - Universidad Nacional de Córdobafalse
dc.title.none.fl_str_mv Comparison of the Accuracy of Clinical Measurements and Cone-Beam Computed Tomography for Assessing Peri-Implant Buccal Bone Thickness Around Titanium and Zirconia Implants
title Comparison of the Accuracy of Clinical Measurements and Cone-Beam Computed Tomography for Assessing Peri-Implant Buccal Bone Thickness Around Titanium and Zirconia Implants
spellingShingle Comparison of the Accuracy of Clinical Measurements and Cone-Beam Computed Tomography for Assessing Peri-Implant Buccal Bone Thickness Around Titanium and Zirconia Implants
Bono, Alejandra
Dental Implants
Zirconia
Cbct
Vestibular Table
title_short Comparison of the Accuracy of Clinical Measurements and Cone-Beam Computed Tomography for Assessing Peri-Implant Buccal Bone Thickness Around Titanium and Zirconia Implants
title_full Comparison of the Accuracy of Clinical Measurements and Cone-Beam Computed Tomography for Assessing Peri-Implant Buccal Bone Thickness Around Titanium and Zirconia Implants
title_fullStr Comparison of the Accuracy of Clinical Measurements and Cone-Beam Computed Tomography for Assessing Peri-Implant Buccal Bone Thickness Around Titanium and Zirconia Implants
title_full_unstemmed Comparison of the Accuracy of Clinical Measurements and Cone-Beam Computed Tomography for Assessing Peri-Implant Buccal Bone Thickness Around Titanium and Zirconia Implants
title_sort Comparison of the Accuracy of Clinical Measurements and Cone-Beam Computed Tomography for Assessing Peri-Implant Buccal Bone Thickness Around Titanium and Zirconia Implants
dc.creator.none.fl_str_mv Bono, Alejandra
Grosso Rossi, Sacha
Ibañez, Juan Carlos
author Bono, Alejandra
author_facet Bono, Alejandra
Grosso Rossi, Sacha
Ibañez, Juan Carlos
author_role author
author2 Grosso Rossi, Sacha
Ibañez, Juan Carlos
author2_role author
author
dc.contributor.none.fl_str_mv https://orcid.org/0000-0003-3684-1947
https://orcid.org/0009-0004-5292-1714
https://orcid.org/0000-0001-6857-4818
dc.subject.none.fl_str_mv Dental Implants
Zirconia
Cbct
Vestibular Table
topic Dental Implants
Zirconia
Cbct
Vestibular Table
dc.description.none.fl_txt_mv Fil: Bono, Alejandra, Liliana. Universidad Nacional de Córdoba. Facultad de Odontología. Departamento de patología oral; Argentina.
Fil: Ibañez, Juan Carlos. Universidad Católica de Córdoba. Facultad de Ciencias de la Salud. Especialización en Implantología Oral; Argentina.
Fil: Grosso Rossi, Sacha. Universidad Nacional de Córdoba. Facultad de Odontología. Cátedra Cirugía II; Argentina.
Dental implants have been an effective solution for restoring oral function and aesthetics in patients with tooth loss. Currently, technological advances have introduced new materials, such as zirconia, which have sparked significant interest in the scientific community. This research aims to compare zirconia and titanium implants and highlight the importance of cone beam computed tomography (CBCT) in the diagnosis of bone complications such as buccal bone dehiscence or fenestrations. Dental implants are a growing part of the prosthetic field. Zirconia has gained popularity as an implant material due to its color and advantageous biological potential. Although it demonstrates soft tissue adhesion and osseointegration potential comparable to commonly used titanium, buccal bone thickness is typically assessed during implant installation or second-stage surgery in submerged implant protocols, but is rarely evaluated thereafter, as conventional oral radiographic diagnosis only assesses interdental marginal bone levels. This factor, however, can have important implications for the long-term stability of peri-implant hard and soft tissues. Computed tomography (CBCT) has significantly expanded the use of CT imaging in implant dentistry due to its potentially reduced radiation dose and cost. However, CBCT images of peri-implant bone can be affected by artifacts that can generate incorrect bone gray values and, consequently, erroneous reconstructions. This study aims to evaluate the accuracy of measuring peri-implant buccal bone using cone-beam computed tomography devices on zirconium implants compared to titanium implants. Objective: To evaluate the accuracy of cone-beam computed tomography on zirconium and titanium implants for measuring buccal peri-implant bone. Implants were placed at different distances from the cortical bone in one prepared porcine rib. The rib will be scanned in a reproducible position using a CT scanner. An examiner will take four measurements on the eight implants: cortical bone thickness from the outer surface of the implant threads at 0, 3, 6, and 9 mm from the top of the implant.
info:eu-repo/semantics/publishedVersion
Fil: Bono, Alejandra, Liliana. Universidad Nacional de Córdoba. Facultad de Odontología. Departamento de patología oral; Argentina.
Fil: Ibañez, Juan Carlos. Universidad Católica de Córdoba. Facultad de Ciencias de la Salud. Especialización en Implantología Oral; Argentina.
Fil: Grosso Rossi, Sacha. Universidad Nacional de Córdoba. Facultad de Odontología. Cátedra Cirugía II; Argentina.
description Fil: Bono, Alejandra, Liliana. Universidad Nacional de Córdoba. Facultad de Odontología. Departamento de patología oral; Argentina.
publishDate 2026
dc.date.none.fl_str_mv 2026-08-01
dc.type.none.fl_str_mv info:eu-repo/semantics/publishedVersion
info:eu-repo/semantics/article
http://purl.org/coar/resource_type/c_6501
info:ar-repo/semantics/articulo
status_str publishedVersion
format article
dc.identifier.none.fl_str_mv Grosso Rossi S., Ibanez J.C. & Bono A. (2026). Comparison of the Accuracy of Clinical Measurements and Cone-Beam Computed Tomography for Assessing Peri-Implant Buccal Bone Thickness Around Titanium and Zirconia Implants. J Oral Dent Health Res, 8(3): 211. https://doi.org/10.36266/JODHR/211
2694-6211
http://hdl.handle.net/11086/562556
10.36266/JODHR/211
identifier_str_mv Grosso Rossi S., Ibanez J.C. & Bono A. (2026). Comparison of the Accuracy of Clinical Measurements and Cone-Beam Computed Tomography for Assessing Peri-Implant Buccal Bone Thickness Around Titanium and Zirconia Implants. J Oral Dent Health Res, 8(3): 211. https://doi.org/10.36266/JODHR/211
2694-6211
10.36266/JODHR/211
url http://hdl.handle.net/11086/562556
dc.language.none.fl_str_mv eng
language eng
dc.rights.none.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.source.none.fl_str_mv reponame:Repositorio Digital Universitario (UNC)
instname:Universidad Nacional de Córdoba
instacron:UNC
reponame_str Repositorio Digital Universitario (UNC)
collection Repositorio Digital Universitario (UNC)
instname_str Universidad Nacional de Córdoba
instacron_str UNC
institution UNC
repository.name.fl_str_mv Repositorio Digital Universitario (UNC) - Universidad Nacional de Córdoba
repository.mail.fl_str_mv oca.unc@gmail.com
_version_ 1877859133850386432
score 12.854657