Practical design of flow meter for mechanical ventilation equipment

Autores
Dugarte, Nelson; Alvarez, Antonio; Dugarte, Edison; Alvarado, Negman; Bhaskar, Sonu
Año de publicación
2021
Idioma
inglés
Tipo de recurso
artículo
Estado
versión aceptada
Descripción
This paper introduces a practical technique for the design of an instrument used in air flow measurement or flowmeter. This instrument is an essential component in the hospital medical ventilation equipment functioning, therefore, the parameters design presented in this article focus on this purpose. However, this instrument can be employed to any measurement scale. The technique is based on indirect flow measurement, using a sensor that converts the flow parameter into a differential pressure measurement. An electronic transducer allows the differential pressure values to be obtained as an electrical signal, which is then digitized and analyzed to obtain the original parameter. The experimental procedure presented in this paper utilizes a computational algorithm to perform the signal analysis; however, given the simplicity of the procedure, this could be adapted to any digital processing card or platform, to show the measurement obtained immediately. Preliminary analyses demonstrated instrument efficiency with sensitivity of 0.0681 liters per second (L/s). Accuracy evaluation showed an average measurement error lesser than 1.4%, with a standard deviation of 0.0612 and normal distribution over the set of test measurements.
Fil: Universidad Tecnológica Nacional. Facultad Regional Mendoza, Argentina
Peer Reviewed
Fuente
Journal of Computer Science & Technology 21(1), 42-48 . (2021)
Materia
Medición de flujo de aire, Instrumentación médica, Tubo Venturi, Equipo de ventilación médica.
Nivel de accesibilidad
acceso abierto
Condiciones de uso
2024-06-24T14:21:09Z
Repositorio
Repositorio Institucional Abierto (UTN)
Institución
Universidad Tecnológica Nacional
OAI Identificador
oai:ria.utn.edu.ar:20.500.12272/11007

id RIAUTN_94825ebeb4b02d36423caf627e4214d0
oai_identifier_str oai:ria.utn.edu.ar:20.500.12272/11007
network_acronym_str RIAUTN
repository_id_str a
network_name_str Repositorio Institucional Abierto (UTN)
spelling Practical design of flow meter for mechanical ventilation equipmentDugarte, NelsonAlvarez, AntonioDugarte, EdisonAlvarado, NegmanBhaskar, SonuMedición de flujo de aire, Instrumentación médica, Tubo Venturi, Equipo de ventilación médica.This paper introduces a practical technique for the design of an instrument used in air flow measurement or flowmeter. This instrument is an essential component in the hospital medical ventilation equipment functioning, therefore, the parameters design presented in this article focus on this purpose. However, this instrument can be employed to any measurement scale. The technique is based on indirect flow measurement, using a sensor that converts the flow parameter into a differential pressure measurement. An electronic transducer allows the differential pressure values to be obtained as an electrical signal, which is then digitized and analyzed to obtain the original parameter. The experimental procedure presented in this paper utilizes a computational algorithm to perform the signal analysis; however, given the simplicity of the procedure, this could be adapted to any digital processing card or platform, to show the measurement obtained immediately. Preliminary analyses demonstrated instrument efficiency with sensitivity of 0.0681 liters per second (L/s). Accuracy evaluation showed an average measurement error lesser than 1.4%, with a standard deviation of 0.0612 and normal distribution over the set of test measurements.Fil: Universidad Tecnológica Nacional. Facultad Regional Mendoza, ArgentinaPeer Reviewed2024-06-24T14:21:09Z2024-06-24T14:21:09Z2021-04-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/acceptedVersionhttp://purl.org/coar/resource_type/c_6501info:ar-repo/semantics/articulopdfapplication/pdfJournal of Computer Science & Technologyhttp://hdl.handle.net/20.500.12272/1100710.24215/16666038.21.e05Journal of Computer Science & Technology 21(1), 42-48 . (2021)reponame:Repositorio Institucional Abierto (UTN)instname:Universidad Tecnológica Nacionalenginfo:eu-repo/semantics/openAccess2024-06-24T14:21:09Zhttp://creativecommons.org/publicdomain/zero/1.0/CC0 1.0 UniversalUniversidad Tecnológica Nacional. Facultad Regional MendozaCC BY (Autoría)2026-10-01T11:57:21Zoai:ria.utn.edu.ar:20.500.12272/11007instacron:UTNInstitucionalhttp://ria.utn.edu.ar/Universidad públicaNo correspondehttp://ria.utn.edu.ar/oaigestionria@rec.utn.edu.ar; fsuarez@rec.utn.edu.arArgentinaNo correspondeNo correspondeNo correspondeopendoar:a2026-10-01 11:57:22.292Repositorio Institucional Abierto (UTN) - Universidad Tecnológica Nacionalfalse
dc.title.none.fl_str_mv Practical design of flow meter for mechanical ventilation equipment
title Practical design of flow meter for mechanical ventilation equipment
spellingShingle Practical design of flow meter for mechanical ventilation equipment
Dugarte, Nelson
Medición de flujo de aire, Instrumentación médica, Tubo Venturi, Equipo de ventilación médica.
title_short Practical design of flow meter for mechanical ventilation equipment
title_full Practical design of flow meter for mechanical ventilation equipment
title_fullStr Practical design of flow meter for mechanical ventilation equipment
title_full_unstemmed Practical design of flow meter for mechanical ventilation equipment
title_sort Practical design of flow meter for mechanical ventilation equipment
dc.creator.none.fl_str_mv Dugarte, Nelson
Alvarez, Antonio
Dugarte, Edison
Alvarado, Negman
Bhaskar, Sonu
author Dugarte, Nelson
author_facet Dugarte, Nelson
Alvarez, Antonio
Dugarte, Edison
Alvarado, Negman
Bhaskar, Sonu
author_role author
author2 Alvarez, Antonio
Dugarte, Edison
Alvarado, Negman
Bhaskar, Sonu
author2_role author
author
author
author
dc.subject.none.fl_str_mv Medición de flujo de aire, Instrumentación médica, Tubo Venturi, Equipo de ventilación médica.
topic Medición de flujo de aire, Instrumentación médica, Tubo Venturi, Equipo de ventilación médica.
dc.description.none.fl_txt_mv This paper introduces a practical technique for the design of an instrument used in air flow measurement or flowmeter. This instrument is an essential component in the hospital medical ventilation equipment functioning, therefore, the parameters design presented in this article focus on this purpose. However, this instrument can be employed to any measurement scale. The technique is based on indirect flow measurement, using a sensor that converts the flow parameter into a differential pressure measurement. An electronic transducer allows the differential pressure values to be obtained as an electrical signal, which is then digitized and analyzed to obtain the original parameter. The experimental procedure presented in this paper utilizes a computational algorithm to perform the signal analysis; however, given the simplicity of the procedure, this could be adapted to any digital processing card or platform, to show the measurement obtained immediately. Preliminary analyses demonstrated instrument efficiency with sensitivity of 0.0681 liters per second (L/s). Accuracy evaluation showed an average measurement error lesser than 1.4%, with a standard deviation of 0.0612 and normal distribution over the set of test measurements.
Fil: Universidad Tecnológica Nacional. Facultad Regional Mendoza, Argentina
Peer Reviewed
description This paper introduces a practical technique for the design of an instrument used in air flow measurement or flowmeter. This instrument is an essential component in the hospital medical ventilation equipment functioning, therefore, the parameters design presented in this article focus on this purpose. However, this instrument can be employed to any measurement scale. The technique is based on indirect flow measurement, using a sensor that converts the flow parameter into a differential pressure measurement. An electronic transducer allows the differential pressure values to be obtained as an electrical signal, which is then digitized and analyzed to obtain the original parameter. The experimental procedure presented in this paper utilizes a computational algorithm to perform the signal analysis; however, given the simplicity of the procedure, this could be adapted to any digital processing card or platform, to show the measurement obtained immediately. Preliminary analyses demonstrated instrument efficiency with sensitivity of 0.0681 liters per second (L/s). Accuracy evaluation showed an average measurement error lesser than 1.4%, with a standard deviation of 0.0612 and normal distribution over the set of test measurements.
publishDate 2021
dc.date.none.fl_str_mv 2021-04-01
2024-06-24T14:21:09Z
2024-06-24T14:21:09Z
dc.type.none.fl_str_mv info:eu-repo/semantics/article
info:eu-repo/semantics/acceptedVersion
http://purl.org/coar/resource_type/c_6501
info:ar-repo/semantics/articulo
format article
status_str acceptedVersion
dc.identifier.none.fl_str_mv Journal of Computer Science & Technology
http://hdl.handle.net/20.500.12272/11007
10.24215/16666038.21.e05
identifier_str_mv Journal of Computer Science & Technology
10.24215/16666038.21.e05
url http://hdl.handle.net/20.500.12272/11007
dc.language.none.fl_str_mv eng
language eng
dc.rights.none.fl_str_mv info:eu-repo/semantics/openAccess
2024-06-24T14:21:09Z
http://creativecommons.org/publicdomain/zero/1.0/
CC0 1.0 Universal
Universidad Tecnológica Nacional. Facultad Regional Mendoza
CC BY (Autoría)
eu_rights_str_mv openAccess
rights_invalid_str_mv 2024-06-24T14:21:09Z
http://creativecommons.org/publicdomain/zero/1.0/
CC0 1.0 Universal
Universidad Tecnológica Nacional. Facultad Regional Mendoza
CC BY (Autoría)
dc.format.none.fl_str_mv pdf
application/pdf
dc.source.none.fl_str_mv Journal of Computer Science & Technology 21(1), 42-48 . (2021)
reponame:Repositorio Institucional Abierto (UTN)
instname:Universidad Tecnológica Nacional
reponame_str Repositorio Institucional Abierto (UTN)
collection Repositorio Institucional Abierto (UTN)
instname_str Universidad Tecnológica Nacional
repository.name.fl_str_mv Repositorio Institucional Abierto (UTN) - Universidad Tecnológica Nacional
repository.mail.fl_str_mv gestionria@rec.utn.edu.ar; fsuarez@rec.utn.edu.ar
_version_ 1877862362704248832
score 12.854657