Practical design of flow meter for mechanical ventilation equipment
- Autores
- Dugarte, Nelson; Alvarez, Antonio; Dugarte, Edison; Alvarado, Negman; Bhaskar, Sonu
- Año de publicación
- 2021
- Idioma
- inglés
- Tipo de recurso
- artículo
- Estado
- versión aceptada
- Descripción
- This paper introduces a practical technique for the design of an instrument used in air flow measurement or flowmeter. This instrument is an essential component in the hospital medical ventilation equipment functioning, therefore, the parameters design presented in this article focus on this purpose. However, this instrument can be employed to any measurement scale. The technique is based on indirect flow measurement, using a sensor that converts the flow parameter into a differential pressure measurement. An electronic transducer allows the differential pressure values to be obtained as an electrical signal, which is then digitized and analyzed to obtain the original parameter. The experimental procedure presented in this paper utilizes a computational algorithm to perform the signal analysis; however, given the simplicity of the procedure, this could be adapted to any digital processing card or platform, to show the measurement obtained immediately. Preliminary analyses demonstrated instrument efficiency with sensitivity of 0.0681 liters per second (L/s). Accuracy evaluation showed an average measurement error lesser than 1.4%, with a standard deviation of 0.0612 and normal distribution over the set of test measurements.
Fil: Universidad Tecnológica Nacional. Facultad Regional Mendoza, Argentina
Peer Reviewed - Fuente
- Journal of Computer Science & Technology 21(1), 42-48 . (2021)
- Materia
- Medición de flujo de aire, Instrumentación médica, Tubo Venturi, Equipo de ventilación médica.
- Nivel de accesibilidad
- acceso abierto
- Condiciones de uso
- 2024-06-24T14:21:09Z
- Repositorio
.jpg)
- Institución
- Universidad Tecnológica Nacional
- OAI Identificador
- oai:ria.utn.edu.ar:20.500.12272/11007
Ver los metadatos del registro completo
| id |
RIAUTN_94825ebeb4b02d36423caf627e4214d0 |
|---|---|
| oai_identifier_str |
oai:ria.utn.edu.ar:20.500.12272/11007 |
| network_acronym_str |
RIAUTN |
| repository_id_str |
a |
| network_name_str |
Repositorio Institucional Abierto (UTN) |
| spelling |
Practical design of flow meter for mechanical ventilation equipmentDugarte, NelsonAlvarez, AntonioDugarte, EdisonAlvarado, NegmanBhaskar, SonuMedición de flujo de aire, Instrumentación médica, Tubo Venturi, Equipo de ventilación médica.This paper introduces a practical technique for the design of an instrument used in air flow measurement or flowmeter. This instrument is an essential component in the hospital medical ventilation equipment functioning, therefore, the parameters design presented in this article focus on this purpose. However, this instrument can be employed to any measurement scale. The technique is based on indirect flow measurement, using a sensor that converts the flow parameter into a differential pressure measurement. An electronic transducer allows the differential pressure values to be obtained as an electrical signal, which is then digitized and analyzed to obtain the original parameter. The experimental procedure presented in this paper utilizes a computational algorithm to perform the signal analysis; however, given the simplicity of the procedure, this could be adapted to any digital processing card or platform, to show the measurement obtained immediately. Preliminary analyses demonstrated instrument efficiency with sensitivity of 0.0681 liters per second (L/s). Accuracy evaluation showed an average measurement error lesser than 1.4%, with a standard deviation of 0.0612 and normal distribution over the set of test measurements.Fil: Universidad Tecnológica Nacional. Facultad Regional Mendoza, ArgentinaPeer Reviewed2024-06-24T14:21:09Z2024-06-24T14:21:09Z2021-04-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/acceptedVersionhttp://purl.org/coar/resource_type/c_6501info:ar-repo/semantics/articulopdfapplication/pdfJournal of Computer Science & Technologyhttp://hdl.handle.net/20.500.12272/1100710.24215/16666038.21.e05Journal of Computer Science & Technology 21(1), 42-48 . (2021)reponame:Repositorio Institucional Abierto (UTN)instname:Universidad Tecnológica Nacionalenginfo:eu-repo/semantics/openAccess2024-06-24T14:21:09Zhttp://creativecommons.org/publicdomain/zero/1.0/CC0 1.0 UniversalUniversidad Tecnológica Nacional. Facultad Regional MendozaCC BY (Autoría)2026-10-01T11:57:21Zoai:ria.utn.edu.ar:20.500.12272/11007instacron:UTNInstitucionalhttp://ria.utn.edu.ar/Universidad públicaNo correspondehttp://ria.utn.edu.ar/oaigestionria@rec.utn.edu.ar; fsuarez@rec.utn.edu.arArgentinaNo correspondeNo correspondeNo correspondeopendoar:a2026-10-01 11:57:22.292Repositorio Institucional Abierto (UTN) - Universidad Tecnológica Nacionalfalse |
| dc.title.none.fl_str_mv |
Practical design of flow meter for mechanical ventilation equipment |
| title |
Practical design of flow meter for mechanical ventilation equipment |
| spellingShingle |
Practical design of flow meter for mechanical ventilation equipment Dugarte, Nelson Medición de flujo de aire, Instrumentación médica, Tubo Venturi, Equipo de ventilación médica. |
| title_short |
Practical design of flow meter for mechanical ventilation equipment |
| title_full |
Practical design of flow meter for mechanical ventilation equipment |
| title_fullStr |
Practical design of flow meter for mechanical ventilation equipment |
| title_full_unstemmed |
Practical design of flow meter for mechanical ventilation equipment |
| title_sort |
Practical design of flow meter for mechanical ventilation equipment |
| dc.creator.none.fl_str_mv |
Dugarte, Nelson Alvarez, Antonio Dugarte, Edison Alvarado, Negman Bhaskar, Sonu |
| author |
Dugarte, Nelson |
| author_facet |
Dugarte, Nelson Alvarez, Antonio Dugarte, Edison Alvarado, Negman Bhaskar, Sonu |
| author_role |
author |
| author2 |
Alvarez, Antonio Dugarte, Edison Alvarado, Negman Bhaskar, Sonu |
| author2_role |
author author author author |
| dc.subject.none.fl_str_mv |
Medición de flujo de aire, Instrumentación médica, Tubo Venturi, Equipo de ventilación médica. |
| topic |
Medición de flujo de aire, Instrumentación médica, Tubo Venturi, Equipo de ventilación médica. |
| dc.description.none.fl_txt_mv |
This paper introduces a practical technique for the design of an instrument used in air flow measurement or flowmeter. This instrument is an essential component in the hospital medical ventilation equipment functioning, therefore, the parameters design presented in this article focus on this purpose. However, this instrument can be employed to any measurement scale. The technique is based on indirect flow measurement, using a sensor that converts the flow parameter into a differential pressure measurement. An electronic transducer allows the differential pressure values to be obtained as an electrical signal, which is then digitized and analyzed to obtain the original parameter. The experimental procedure presented in this paper utilizes a computational algorithm to perform the signal analysis; however, given the simplicity of the procedure, this could be adapted to any digital processing card or platform, to show the measurement obtained immediately. Preliminary analyses demonstrated instrument efficiency with sensitivity of 0.0681 liters per second (L/s). Accuracy evaluation showed an average measurement error lesser than 1.4%, with a standard deviation of 0.0612 and normal distribution over the set of test measurements. Fil: Universidad Tecnológica Nacional. Facultad Regional Mendoza, Argentina Peer Reviewed |
| description |
This paper introduces a practical technique for the design of an instrument used in air flow measurement or flowmeter. This instrument is an essential component in the hospital medical ventilation equipment functioning, therefore, the parameters design presented in this article focus on this purpose. However, this instrument can be employed to any measurement scale. The technique is based on indirect flow measurement, using a sensor that converts the flow parameter into a differential pressure measurement. An electronic transducer allows the differential pressure values to be obtained as an electrical signal, which is then digitized and analyzed to obtain the original parameter. The experimental procedure presented in this paper utilizes a computational algorithm to perform the signal analysis; however, given the simplicity of the procedure, this could be adapted to any digital processing card or platform, to show the measurement obtained immediately. Preliminary analyses demonstrated instrument efficiency with sensitivity of 0.0681 liters per second (L/s). Accuracy evaluation showed an average measurement error lesser than 1.4%, with a standard deviation of 0.0612 and normal distribution over the set of test measurements. |
| publishDate |
2021 |
| dc.date.none.fl_str_mv |
2021-04-01 2024-06-24T14:21:09Z 2024-06-24T14:21:09Z |
| dc.type.none.fl_str_mv |
info:eu-repo/semantics/article info:eu-repo/semantics/acceptedVersion http://purl.org/coar/resource_type/c_6501 info:ar-repo/semantics/articulo |
| format |
article |
| status_str |
acceptedVersion |
| dc.identifier.none.fl_str_mv |
Journal of Computer Science & Technology http://hdl.handle.net/20.500.12272/11007 10.24215/16666038.21.e05 |
| identifier_str_mv |
Journal of Computer Science & Technology 10.24215/16666038.21.e05 |
| url |
http://hdl.handle.net/20.500.12272/11007 |
| dc.language.none.fl_str_mv |
eng |
| language |
eng |
| dc.rights.none.fl_str_mv |
info:eu-repo/semantics/openAccess 2024-06-24T14:21:09Z http://creativecommons.org/publicdomain/zero/1.0/ CC0 1.0 Universal Universidad Tecnológica Nacional. Facultad Regional Mendoza CC BY (Autoría) |
| eu_rights_str_mv |
openAccess |
| rights_invalid_str_mv |
2024-06-24T14:21:09Z http://creativecommons.org/publicdomain/zero/1.0/ CC0 1.0 Universal Universidad Tecnológica Nacional. Facultad Regional Mendoza CC BY (Autoría) |
| dc.format.none.fl_str_mv |
pdf application/pdf |
| dc.source.none.fl_str_mv |
Journal of Computer Science & Technology 21(1), 42-48 . (2021) reponame:Repositorio Institucional Abierto (UTN) instname:Universidad Tecnológica Nacional |
| reponame_str |
Repositorio Institucional Abierto (UTN) |
| collection |
Repositorio Institucional Abierto (UTN) |
| instname_str |
Universidad Tecnológica Nacional |
| repository.name.fl_str_mv |
Repositorio Institucional Abierto (UTN) - Universidad Tecnológica Nacional |
| repository.mail.fl_str_mv |
gestionria@rec.utn.edu.ar; fsuarez@rec.utn.edu.ar |
| _version_ |
1877862362704248832 |
| score |
12.854657 |