ECG Multilead QT interval estimation using support vector machines
- Autores
- Dugarte, Nelson
- Año de publicación
- 2019
- Idioma
- inglés
- Tipo de recurso
- artículo
- Estado
- versión aceptada
- Descripción
- This work reports a multilead QT interval measurement algorithm for a high-resolution digital electrocardiograph. The software enables off-line ECG processing including QRS detection as well as an accurate multilead QT interval detection algorithm using support vector machines (SVMs). Two fiducial points ( and ) are estimated using the SVM algorithm on each incoming beat. This enables segmentation of the current beat for obtaining the P, QRS, and T waves. The QT interval is estimated by updating the QT interval on each lead, considering shifting techniques with respect to a valid beat template. The validation of the QT interval measurement algorithm is attained using the Physionet PTB diagnostic ECG database showing a percent error of with respect to the database annotations. The usefulness of this software tool is also tested by considering the analysis of the ECG signals for a group of 60 patients acquired using our digital electrocardiograph. In this case, the validation is performed by comparing the estimated QT interval with respect to the estimation obtained using the Cardiosoft software providing a percent error of .
Fil: Universidad Tecnológica Nacional. Facultad Regional Mendoza, Argentina
Peer Reviewed - Materia
- Vector machines, Interval estimation, Multilead QT
- Nivel de accesibilidad
- acceso abierto
- Condiciones de uso
- 2024-06-25T13:21:35Z
- Repositorio
.jpg)
- Institución
- Universidad Tecnológica Nacional
- OAI Identificador
- oai:ria.utn.edu.ar:20.500.12272/11039
Ver los metadatos del registro completo
| id |
RIAUTN_c646b2d508d69017b5b327439e55a857 |
|---|---|
| oai_identifier_str |
oai:ria.utn.edu.ar:20.500.12272/11039 |
| network_acronym_str |
RIAUTN |
| repository_id_str |
a |
| network_name_str |
Repositorio Institucional Abierto (UTN) |
| spelling |
ECG Multilead QT interval estimation using support vector machinesDugarte, NelsonVector machines, Interval estimation, Multilead QTThis work reports a multilead QT interval measurement algorithm for a high-resolution digital electrocardiograph. The software enables off-line ECG processing including QRS detection as well as an accurate multilead QT interval detection algorithm using support vector machines (SVMs). Two fiducial points ( and ) are estimated using the SVM algorithm on each incoming beat. This enables segmentation of the current beat for obtaining the P, QRS, and T waves. The QT interval is estimated by updating the QT interval on each lead, considering shifting techniques with respect to a valid beat template. The validation of the QT interval measurement algorithm is attained using the Physionet PTB diagnostic ECG database showing a percent error of with respect to the database annotations. The usefulness of this software tool is also tested by considering the analysis of the ECG signals for a group of 60 patients acquired using our digital electrocardiograph. In this case, the validation is performed by comparing the estimated QT interval with respect to the estimation obtained using the Cardiosoft software providing a percent error of .Fil: Universidad Tecnológica Nacional. Facultad Regional Mendoza, ArgentinaPeer Reviewed2024-06-25T13:21:35Z2024-06-25T13:21:35Z2019-01-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/acceptedVersionhttp://purl.org/coar/resource_type/c_6501info:ar-repo/semantics/articulopdfapplication/pdfJournal of Healthcare Engineeringhttp://hdl.handle.net/20.500.12272/11039https://doi.org/10.1155/2019/6371871enginfo:eu-repo/semantics/openAccess2024-06-25T13:21:35Zhttp://creativecommons.org/publicdomain/zero/1.0/CC0 1.0 UniversalUniversidad Tecnológica Nacional. Facultad Regional MendozaCC BY (Autoría)reponame:Repositorio Institucional Abierto (UTN)instname:Universidad Tecnológica Nacional2026-10-29T11:27:46Zoai:ria.utn.edu.ar:20.500.12272/11039instacron:UTNInstitucionalhttp://ria.utn.edu.ar/Universidad públicaNo correspondehttp://ria.utn.edu.ar/oaigestionria@rec.utn.edu.ar; fsuarez@rec.utn.edu.arArgentinaNo correspondeNo correspondeNo correspondeopendoar:a2026-10-29 11:27:46.567Repositorio Institucional Abierto (UTN) - Universidad Tecnológica Nacionalfalse |
| dc.title.none.fl_str_mv |
ECG Multilead QT interval estimation using support vector machines |
| title |
ECG Multilead QT interval estimation using support vector machines |
| spellingShingle |
ECG Multilead QT interval estimation using support vector machines Dugarte, Nelson Vector machines, Interval estimation, Multilead QT |
| title_short |
ECG Multilead QT interval estimation using support vector machines |
| title_full |
ECG Multilead QT interval estimation using support vector machines |
| title_fullStr |
ECG Multilead QT interval estimation using support vector machines |
| title_full_unstemmed |
ECG Multilead QT interval estimation using support vector machines |
| title_sort |
ECG Multilead QT interval estimation using support vector machines |
| dc.creator.none.fl_str_mv |
Dugarte, Nelson |
| author |
Dugarte, Nelson |
| author_facet |
Dugarte, Nelson |
| author_role |
author |
| dc.subject.none.fl_str_mv |
Vector machines, Interval estimation, Multilead QT |
| topic |
Vector machines, Interval estimation, Multilead QT |
| dc.description.none.fl_txt_mv |
This work reports a multilead QT interval measurement algorithm for a high-resolution digital electrocardiograph. The software enables off-line ECG processing including QRS detection as well as an accurate multilead QT interval detection algorithm using support vector machines (SVMs). Two fiducial points ( and ) are estimated using the SVM algorithm on each incoming beat. This enables segmentation of the current beat for obtaining the P, QRS, and T waves. The QT interval is estimated by updating the QT interval on each lead, considering shifting techniques with respect to a valid beat template. The validation of the QT interval measurement algorithm is attained using the Physionet PTB diagnostic ECG database showing a percent error of with respect to the database annotations. The usefulness of this software tool is also tested by considering the analysis of the ECG signals for a group of 60 patients acquired using our digital electrocardiograph. In this case, the validation is performed by comparing the estimated QT interval with respect to the estimation obtained using the Cardiosoft software providing a percent error of . Fil: Universidad Tecnológica Nacional. Facultad Regional Mendoza, Argentina Peer Reviewed |
| description |
This work reports a multilead QT interval measurement algorithm for a high-resolution digital electrocardiograph. The software enables off-line ECG processing including QRS detection as well as an accurate multilead QT interval detection algorithm using support vector machines (SVMs). Two fiducial points ( and ) are estimated using the SVM algorithm on each incoming beat. This enables segmentation of the current beat for obtaining the P, QRS, and T waves. The QT interval is estimated by updating the QT interval on each lead, considering shifting techniques with respect to a valid beat template. The validation of the QT interval measurement algorithm is attained using the Physionet PTB diagnostic ECG database showing a percent error of with respect to the database annotations. The usefulness of this software tool is also tested by considering the analysis of the ECG signals for a group of 60 patients acquired using our digital electrocardiograph. In this case, the validation is performed by comparing the estimated QT interval with respect to the estimation obtained using the Cardiosoft software providing a percent error of . |
| publishDate |
2019 |
| dc.date.none.fl_str_mv |
2019-01-01 2024-06-25T13:21:35Z 2024-06-25T13:21:35Z |
| dc.type.none.fl_str_mv |
info:eu-repo/semantics/article info:eu-repo/semantics/acceptedVersion http://purl.org/coar/resource_type/c_6501 info:ar-repo/semantics/articulo |
| format |
article |
| status_str |
acceptedVersion |
| dc.identifier.none.fl_str_mv |
Journal of Healthcare Engineering http://hdl.handle.net/20.500.12272/11039 https://doi.org/10.1155/2019/6371871 |
| identifier_str_mv |
Journal of Healthcare Engineering |
| url |
http://hdl.handle.net/20.500.12272/11039 https://doi.org/10.1155/2019/6371871 |
| dc.language.none.fl_str_mv |
eng |
| language |
eng |
| dc.rights.none.fl_str_mv |
info:eu-repo/semantics/openAccess 2024-06-25T13:21:35Z http://creativecommons.org/publicdomain/zero/1.0/ CC0 1.0 Universal Universidad Tecnológica Nacional. Facultad Regional Mendoza CC BY (Autoría) |
| eu_rights_str_mv |
openAccess |
| rights_invalid_str_mv |
2024-06-25T13:21:35Z http://creativecommons.org/publicdomain/zero/1.0/ CC0 1.0 Universal Universidad Tecnológica Nacional. Facultad Regional Mendoza CC BY (Autoría) |
| dc.format.none.fl_str_mv |
pdf application/pdf |
| dc.source.none.fl_str_mv |
reponame:Repositorio Institucional Abierto (UTN) instname:Universidad Tecnológica Nacional |
| reponame_str |
Repositorio Institucional Abierto (UTN) |
| collection |
Repositorio Institucional Abierto (UTN) |
| instname_str |
Universidad Tecnológica Nacional |
| repository.name.fl_str_mv |
Repositorio Institucional Abierto (UTN) - Universidad Tecnológica Nacional |
| repository.mail.fl_str_mv |
gestionria@rec.utn.edu.ar; fsuarez@rec.utn.edu.ar |
| _version_ |
1880396329357672448 |
| score |
13.334379 |